Total Mercury Analyzer HM-1400 TR
|
Measuring device for fully-automatic and continuous mercury analysis in smoke gas (without wet chemistry).

|
|
|
Features
|
|
|
 |
Maintenance-free (6 months) dry reactor
|
 |
High operational safety
|
 |
Easy maintenance
|
 |
Low cross sensitivities
|
 |
Easily legible LC display
|
|
|
Applications
|
|
|
 |
Waste incinerations (municipal waste, industrial waste, hospital waste)
|
 |
Sewage sludge incineration
|
 |
Hazardous waste incineration
|
 |
Steel plants with scrap metal preparation
|
 |
Contaminated soil burning plants
|
 |
Crematoriums
|
 |
Mercury mines and refineries
|
 |
Fluorescent light bulb recycling plants
|
|
|
|
|
|
measurements
|
total mercury
|
|
measuring ranges
|
0–45 ... 0–500 µg/Nm3
|
|
measuring principle
|
UV-absorption
|
|
flue gas temperature
|
0–250°C
|
|
flue gas pressure
|
-50 up to +50 hPa
|
|
duct diameter
|
>0.5 m
|
|
ambient temperature
|
+5 up to +30°C
|
|
protection
|
IP40 (IP55)
|
|
measuring outputs
|
2 x 0 / 4–20 mA / 500 Ohm
|
|
digital outputs
|
4 relay outputs, permissable load 250 V / 100 VA
|
|
digital inputs
|
1 potential free input
|
|
accuracy
|
<1% of measuring range
|
|
detection limit
|
<1 µg/Nm3
|
|
reference point drift
|
<2% of measuring range/month
|
|
zero point drift
|
automatic zero point correction
|
|
supply voltage
|
230 VAC, 50 Hz, 1200 VA, sample probe: 650 VA, sample line: 100 VA/m
|
|
dimensions (H x W x D)
|
cabinet 1600 x 800 x 500 mm
|
|
weight
|
220 kg
|
|
purge air supply
|
pressurized air 3–6 bar
|
|
protection
|
IP54
|
More information 
Article in IET 
|
|
|